Impact of Depth and ROI Size on the Reliability of SWE Measurement — МЕДТРЕЙН Asia
Elastography

Impact of Depth and ROI Size on the Reliability of SWE Measurement

Briefly. The reliability of SWE directly depends on the measurement geometry: for 2D-SWE, the ROI size should be at least 10 mm, and the measurement box should be positioned 15–20 mm below the liver capsule. The optimal depth for the push pulse is 4–4.5 cm from the transducer; beyond 6–7 cm, shear wave generation weakens (WFUMB, 2024).

ROI Size in 2D-SWE

According to WFUMB (2024), in 2D-SWE, the ROI (measurement box) should be at least 10 mm. Unlike p-SWE, where the ROI size is fixed, in 2D-SWE it can be adjusted (Ferraioli et al., 2018). The ROI should ideally be placed closer to the center of the field of view (FOV) as measurement errors often occur at the FOV boundaries (Ferraioli et al., 2018).

Depth of ROI Placement

According to WFUMB (2024), the measurement box in liver studies should be positioned 15–20 mm below the capsule. The 2D-SWE field of view can be shifted closer to the capsule provided reverberation artifacts are avoided.

The key physical constraint is the energy distribution of the ARFI push pulse: in most ultrasound systems, the push pulse peak is achieved at a depth of 4–4.5 cm from the transducer, which is the optimal zone for measurements. At a depth of 6–7 cm, the push pulse weakens, limiting adequate shear wave generation (WFUMB, 2024).

The EFSUMB Course Book (Sporea et al., 2018) recommends placing the ROI at a depth of more than 2 cm but not deeper than 8 cm. According to Bamber et al. (2013), shear wave speed can be measured to a depth of approximately 8 cm.

ROI Placement Relative to Structures

The ROI should be placed in an area free of large blood vessels, bile ducts, and mass lesions (WFUMB, 2024; ACR/SRU, 2024; Sporea et al., 2018). Measurements are performed during neutral breath-hold (WFUMB, 2024).

Number of Measurements and Quality Control

ParameterRecommendationSource
Number of p-SWE Measurements10 measurements on 10 independent images in one location, median usedACR/SRU, 2024
Number of 2D-SWE MeasurementsAt least 5 when using quality assessment parameter; minimum 3 (median + IQR)ACR/SRU, 2024; Sporea et al., 2018
IQR/M (kPa)≤30% — acceptable for accurate data setACR/SRU, 2024
IQR/M (m/s)≤15%ACR/SRU, 2024

Operator Training

According to EFSUMB (Dietrich et al., 2017), a novice is recommended to perform at least 50 supervised 2D-SWE measurements to achieve stable results; an experienced operator demonstrates higher reproducibility over time.

Frequently asked questions

What should be the minimum ROI size in 2D-SWE?

At least 10 mm (WFUMB, 2024).

At what depth below the liver capsule should the measurement box be placed?

15–20 mm below the liver capsule (WFUMB, 2024).

What is the optimal measurement depth relative to the transducer?

4–4.5 cm — zone of push pulse peak; at 6–7 cm, shear wave generation weakens (WFUMB, 2024).

What IQR/M threshold is considered acceptable?

≤30% for kPa and ≤15% for m/s (ACR/SRU, 2024).

How many measurements are needed in 2D-SWE?

Minimum 3 (median + IQR) according to EFSUMB; at least 5 when using quality parameter according to ACR/SRU (2024).

The material is intended for specialists and does not replace clinical judgment. Threshold values are periodically reviewed — refer to the current edition of the applicable consensus.
Sources: WFUMB Guideline on Liver Multiparametric Ultrasound – Part 1, Ferraioli et al., 2024; ACR-SRU Practice Parameter, 2024; Ferraioli et al., 2018; Sporea et al., EFSUMB Course Book, 2018; Dietrich et al., EFSUMB 2017; Bamber et al., EFSUMB 2013; Săftoiu et al., EFSUMB 2019.
View specialty courses: Elastography →
Спросить Alex Отвечу на любой вопрос · 24/7 · на любом языке